KROHNE CORIMASS MFC 85 Interface DE Benutzerhandbuch

Seite 34

Advertising
background image

34

6.7

Diagnostikcodes

Diese beziehen sich auf Modbus Befehl 08.

Sub

command

Description

0

Echo Query command

1

Restart communcations

1

2

Return 16bit status register

2

3

Not supported

4

Turn off communications

1

5

Not supported

6

Not supported

7

Not supported

8

Not supported

9

Not supported

10

Clear event log

11

Return bus message count

12

Return CRC error count

13

Return Exception count

14

Return Slave message count

15

Return No response count

16

Return NAK count

17

Return SlaveBusy count

18

Return Communications Overrun
count

1

. Diese Befehle geben keine Antwort

2

. Status Register

Bit No

Meaning

00

Zero Error - Excessive flow measured during zero calibration.

01

Temperature measured > 180 C or < -25 C.

02

Sensor A out of range.

03

Sensor B out of range.

04

Ratio of sensors A/B exception. One sensor reading is >> than other.

05

Sensor A has a large DC offset.

06

Sensor B has a large DC offset.

07

No synchronization with primary head oscillations.

08

ROM checksum error detected on power-up.

09

EEPROM save error.

10

Non Volatile RAM checksum error detected on power-up

11

Power Failure recorded.

12

Watchdog System Reboot has been activated.

13

Software exception registered.

14

Temperature drifted +/- 30 C from zero calibration temperature.

15

Current Loop measurement Out of Range.

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: