Gossen MAVOLUX 5032 C BASE Benutzerhandbuch
Seite 45
Advertising

GOSSEN Foto- und Lichtmesstechnik GmbH
21
Characteristics MAVOLUX 5032C BASE
Measuring Quantity
Measuring Range
Measuring Range
Resolution
Resolution
in Lux (lx)
in footcandle (fc)
in lx
in fc
Illuminance
I
0.1... 199.9
0.01... 19.99
0.1
0.01
II
1... 1 999
0.1... 199.9
1
0.1
III
10... 19 900
1... 1 999
10
1
IV
100... 199 000
10... 19 990
100
10
Most Important Error Limits MAVOLUX 5032 C BASE
Characteristics
Admissible Error according to
DIN 5032 Klasse C
Typical Error
MAVOLUX 5032C BASE
V(
)-Adaptation (f
1
)
9%
7,5%
True Cosine Evaluation (f
2
)
6%
2,0%
Linearity (f
3
)
5%
1,5%
Adjustment Error (f
11
)
2%
1,0%
Overall Error (f
ges
)
20%
15,0%
Advertising