Sony SIR-3400H Benutzerhandbuch
Seite 123

(E)
17-11
SIR-3400H / SIR-3032i / SIR-3032W
Command
Contents
ERR?
Error check
Command format
ERR?
Returned format
ERR : <1><2>...<5>
<1> : Power supply system
0=No error, 1=Error
<2> : DSP system
0=No error, 1=Error
<3> : Analog system
0=No error, 1=Error
<4> : Recording system
0=No error, 1=Error
<5> : Peripherals
0=No error, 1=Error
Related commands : CAL, SLF
FMT?
Recording media formatting execution status
Command format
FMT?/[ ]
Returned format
FMT : <1><2><3><4>
<1> :
0= Other than below (after power-on or STOP command execution, etc.)
1= Recording media formatting underway
2= Recording media formatting finished normally
(becomes “0” after FMT?)
3= Recording media formatting finished abnormally
(becomes “0” after FMT?)
<2><3><4> : ID number (001 to 999)
Related command : FMT
FTP?
Frequency type readout
Command format
FTP?<1>/[ ]
Returned format
FTP : <1>
<1> : 0=48 k sampling
1=65.536 k sampling
IDS?
Recording media ID number readout
Command format
IDS?/[ ]
Returned format
IDS : <1><2><3>
<1><2><3> : ID number (000 to 999)
Related command : IDS
INP?
Input format setting and readout
Command format
INP?<1><2>/[ ]
Returned format
INP : <1><2>...<18>
<1> : 0=Recording unit setting data, 1=Recording media data
<2> : Measuring unit ID number
0=Measuring unit 1
1 to 16 CH
1=Measuring unit 1
17 to 32 CH
2=Measuring unit 2
33 to 48 CH
3=Measuring unit 2
49 to 64 CH
4=Measuring unit 3
65 to 80 CH
5=Measuring unit 3
81 to 96 CH
6=Measuring unit 4
97 to 112 CH
7=Measuring unit 4
113 to 128 CH
<3> : Input format for CH1
0=DIRECT-DC coupling
1=DIRECT-AC coupling
2=IEPE
3=TEDS
<4> : Input format for CH2
<18> : Input format for CH16
Related command : INP
17. Control by Telnet
.
.
.