17-14
(E)
SIR-3400H / SIR-3032i / SIR-3032W
Command
Contents
RNG?
Input range setting value readout
Command format
RNG? <1><2>/[ ]
<1> : 0=Recording unit setting value, 1=Recording media data
<2> : Measuring unit ID number
0=Measuring unit 1
1=Measuring unit 1
2=Measuring unit 2
3=Measuring unit 2
4=Measuring unit 3
5=Measuring unit 3
6=Measuring unit 4
7=Measuring unit 4
Returned format
RNG : <1><2>...<16>
<1> : CH 1 input range code
<2> : CH 2 input range code
<16> : CH 16 input range code
Input range code
0=RSV, 1=10 V, 2=5 V, 3=2 V, 4=1 V, 5=0.5 V, 6=0.2 V, 7=0.1 V
Related command : RNG
RPT?
Auto repeat status readout
Command format
RPT? /[ ]
Returned format
RPT : <1>
<1> : 0=Other than below
1=Auto repeat — recording
2=Auto repeat — rewinding
Related command : RPT
SCH?
Search status
Command format
SCH? /[ ]
Returned format
SCH : <1>
<1> : 0=Other than below (after power-on STOP command execution)
1=Search underway
2=Search finished (FOUND)
3=Search finished (NOT FOUND)
Related command : SCH
SLF?
Diagnostics status
Command format
SLF? /[ ]
Returned format
SLF : <1>
<1> : 0=Other than below (after power-on STOP command execution)
1=Diagnostics underway
2=Diagnostics finished (OK)
3=Diagnostics finished (NG)
Related command : SLF
Related command : ERR?
SPD?
Recording media speed readout
Command format
SPD? <1>/[ ]
<1> : 0=Recording unit setting value, 1=Recording media data
Returned format
SPD : <1>
<1> : 0=1
x, 1=1/2 x, 2=1/4 x, 3=1/8 x, 4=1/16 x, A=x2, B=x4, F=EXT
Related command : SPD
.
.
.