9.3.2
Mastercycler
9.3.3
Thermomixer und ThermoStat
RT = Raumtemperatur
9.3.4
ThermoMixer C, TermoMixer F1.5, ThermoMixer FP und ThermoStat C
Mastercycler Familie
Temperatur
Fehlergrenzen
Mastercycler nexus
35 °C ±0,6 K
95 °C
±1,0 K
Mastercycler ep
35 °C ±0,6 K
95 °C
±1,0 K
Mastercycler pro
35 °C ±0,6 K
95 °C
±1,0 K
Temperatur
Thermomixer
comfort
Thermomixer
compact
ThermoStat plus
0 °C
———
———
±1,0 K
10 °C unter RT
±2,0 K
———
———
37 °C
±0,5 K
±1,0 K
±0,5 K
56 °C
±2,0 K
±2,0 K
±1,0 K
75 °C
±2,0 K
±2,0 K
±1,0 K
90 °C
±2,0 K
±2,0 K
±1,0 K
Temperatur
Fehlergrenzen
20 °C-45 °C
±1,0 K
<20 °C und >45 °C
±1,0 K