2 mastercycler, 3 thermomixer und thermostat – Eppendorf Temperature Verification System Benutzerhandbuch

Seite 35

Advertising
background image

Temperature Verification System USB – Single-channel — Bedienungsanleitung

35

9.3.2

Mastercycler

9.3.3

Thermomixer und ThermoStat

RT = Raumtemperatur

9.3.4

ThermoMixer C, TermoMixer F1.5, ThermoMixer FP und ThermoStat C

Mastercycler Familie

Temperatur

Fehlergrenzen

Mastercycler nexus

35 °C ±0,6 K

95 °C

±1,0 K

Mastercycler ep

35 °C ±0,6 K

95 °C

±1,0 K

Mastercycler pro

35 °C ±0,6 K

95 °C

±1,0 K

Temperatur

Thermomixer

comfort

Thermomixer

compact

ThermoStat plus

0 °C

———

———

±1,0 K

10 °C unter RT

±2,0 K

———

———

37 °C

±0,5 K

±1,0 K

±0,5 K

56 °C

±2,0 K

±2,0 K

±1,0 K

75 °C

±2,0 K

±2,0 K

±1,0 K

90 °C

±2,0 K

±2,0 K

±1,0 K

Temperatur

Fehlergrenzen

20 °C-45 °C

±1,0 K

<20 °C und >45 °C

±1,0 K

Advertising