3 intended use, 4 installation and operation, 1 mounting exchangeable thermoblocks – Eppendorf Thermoblocks for Thermostat with IsoTherm-System Benutzerhandbuch
Seite 6

6
Instructions for use
3
Intended use
The exchangeable thermoblocks and the IsoTherm System are intended exclusively for
indoor use. The exchangeable thermoblocks are used to hold tubes and plates, the
IsoTherm system to cool tubes during sample preparation.
The exchangeable thermoblocks and the IsoTherm System must only be operated by
appropriately qualified and trained specialist personnel. Carefully read the operating
manual before starting work and familiarize yourself with the application.
4
Installation and operation
4.1 Mounting exchangeable thermoblocks
WARNING!
Warning! Injury or property damage from improperly attached
exchangeable thermoblocks.
If the exchangeable thermoblocks are not attached properly, they can become
detached from the device or break.
Tighten the screws of the exchangeable thermoblocks fingertight.
WARNING!
Warning! Contamination due to opening tube cover.
Covers of micro test tubes can snap open involuntarily in the following cases,
causing sample material to escape:
•
high vapor pressure of the content
•
improperly sealed cover
•
damaged sealing lip
Always check that micro test tubes have been sealed tightly before use.
WARNING!
Warning! Risk of burns from hot surfaces.
The exchangeable thermoblock can be very hot after heating and cause
burns.
Allow the heated exchangeable thermoblock to cool down completely
before removing the exchangeable thermoblock.