Installation Manual

tiqa01d2 (2004-09)

Validity

This document applies to products with the part numbers:

250-900

(VCC500, 90 … 250 VAC)

399-552

(VGC012, 120 VAC)

399-553

(VGC012, 230 VAC)

399-554

(VGC012, 100 VAC)

399-556

(VGC023, 120 VAC)

399-557

(VGC023, 230 VAC)

399-558

(VGC023, 100 VAC)

398-010

(VGC401, 90 … 250 VAC)

398-020

(VGC402, 90 … 250 VAC)

398-021

(VGC403, 90 … 250 VAC)

We reserve the right to make technical changes without prior
notice.

Further Information

Details about the intended use, scope of delivery, installation,
operation, etc. you will find in the respective Operating
Manuals:

VCC500

→ sina38e1

VGC012

→ tinb06e1

VGC023

→ tinb05e1

VGC401

→ tinb01e1

VGC402/403

→ tinb07e1

The Operating Manuals you will find on the CD-ROM (in-
cluded in scope of delivery of the control units) or they can be
downloaded from our website.

Safety

Symbols Used

DANGER

Information on preventing any kind of physical injury.

Personnel Qualifications

Skilled personnel

All work described in this document may only be carried out
by persons who have suitable technical training and the
necessary experience or who have been instructed by the
end-user of the product.

Liability and Warranty

INFICON assumes no liability and the warranty becomes null
and void if the end-user or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of interventions (modifications, alterations

etc.) on the product

• use the product with accessories not listed in the cor-

responding product documentation.

General Safety Instructions

Adhere to the applicable regulations and take the necessary
precautions for all work you are going to do and consider the
safety instructions in this document.

DANGER

Caution: mains voltage
Contact with live parts is extremely hazardous
when any objects are introduced or any liquids
penetrate into the unit.
Make sure no objects enter through the louvers
and no liquids penetrate into the equipment.

Communicate the safety instructions to all other users.

Installation

Skilled personnel

The unit may only be installed by persons who
have suitable technical training and the neces-
sary experience or who have been instructed by
the end-user of the product.

DANGER

Caution: damaged product
The product can be damaged. Putting a dam-
aged product into operation can be extremely
hazardous.
In case of visible damages, make sure the
product is not put into operation.

Rack and Control Panel Installation

(

VCC500

,

VGC012

,

VGC023

,

VGC401

,

VGC402, VGC403

)

DANGER

Caution: protection class of the rack or of the
control panel
If the product is installed in a rack or in a control
panel, it is likely to lower the protection class of
the rack or the control panel (protection against
foreign bodies and water) e.g. according to the
EN 60204-1 regulations for switching cabinets.
Take appropriate measures for the rack or for
the control panel to meet the specifications of the
protection class.

The admissible maximum ambient temperature must
not be exceeded (

→ respective Operating Manual).

Fastening screws are supplied the Control Units for installa-
tion in a 19" rack chassis adapter acc. to DIN 41 494.
Required cut-out for a control panel installation

→ respective

Operating Manual.

Use as Desk-Top Unit

(

VGC012

,

VGC023

,

VGC401

,

VGC402, VGC403

)

Select a location where the admissible maximum
ambient temperature (

→ respective Operating

Manual) is not exceeded (e.g. due to sun irradia-
tion).

For the use as desk-top unit, self-adhesive rubber feet as well
as a slip-on rubber bar are supplied with the Control Units.

Mains Power Connection

230 V

?

120 V

?

100 V

?

(VGC012, VGC023)
Changing the line voltage
setting and exchanging the
fuse

→ tinb06e1 and

tinb05e1.

DANGER

Caution: line voltage
Incorrectly grounded products can be extremely
hazardous in the event of a fault.
Use only a 3-conductor power cable with pro-
tective ground. The power connector may only
be plugged into a socket with a protective
ground. The protection must not be nullified by
an extension cable without protective ground.

The unit is supplied with a power cord. If the mains connector
is not compatible with your system, use your own, suitable
cable with protective ground (3

×1.5 mm

2

).

The socket must be

fuse - protected with

10 A max

VCC500

If the unit is installed in a switching cabinet, the mains voltage
should be supplied and turned on via a central distributor.

Connection of Gauges, Interface, …

→ respective Operating Manual

Operation

Turning the Unit On

The power switch is on the rear of the unit. Turn the unit on
with the power switch (or centrally, via a switched power dis-
tributor, if the unit is incorporated in a rack).

Turning the Unit Off

The power switch is on the rear of the unit. Turn the unit off
with the power switch (or centrally, via a switched power dis-
tributor, if the unit is incorporated in a rack).

Wait at least 10 s before turning the unit on again in
order for it to correctly initialize itself.