3 connect device, 4 remove the transport protection pad of the rotor, 5 the device is now ready to use – Eppendorf 5418 R Centrifuge Benutzerhandbuch

Seite 2: 3 packing, 4 passing on the device, 5 disposal

Advertising
background image

Transporting, Installing, Transferring and Disposing of the Centrifuge 5418 R

2

2.3 Connect device

2.4 Remove the transport protection pad of the rotor

2.5 The device is now ready to use.

Retain the packaging and all transport protection pads for shipping the device at a later date.

3 Packing

Pack the centrifuge in reverse order.

4 Passing On the Device

When passing the device on to third parties, be sure to include the operating manual and

these instructions for use.

5 Disposal

Observe the applicable legal regulations when disposing of the device. The disposal is regulated

within the European Community by national regulations based on EU Directive 2002/96/EC on

waste electrical and electronic equipment (WEEE). According to these regulations, any devices

supplied after August 13, 2005 on a business-to-business basis, to which this product is

assigned, may no longer be disposed of in municipal or household waste. For more information,

please contact your supplier.

After installation, wait four hours before

switching the centrifuge on in order to avoid

damage to the compressor.

Check that the mains voltage and frequency

match the requirements on the device type

plate (see rear side of the device) and then

connect the device to the power supply.

1. Switch on the device at the mains power

switch.

2. Open the centrifuge lid using the open key.
3. Turn the rotor nut counterclockwise using the

rotor key supplied.

4. Remove rotor by lifting vertically.
5. Remove transport protection pad.
6. Place rotor vertically onto the motor shaft.
7. Turn the rotor nut with rotor key clockwise

until the rotor nut is tightened.

Advertising