3 connect device, 4 remove the transport protection pad of the rotor, 5 the device is now ready to use – Eppendorf 5418 R Centrifuge Benutzerhandbuch
Seite 2: 3 packing, 4 passing on the device, 5 disposal

Transporting, Installing, Transferring and Disposing of the Centrifuge 5418 R
2
2.3 Connect device
2.4 Remove the transport protection pad of the rotor
2.5 The device is now ready to use.
Retain the packaging and all transport protection pads for shipping the device at a later date.
3 Packing
Pack the centrifuge in reverse order.
4 Passing On the Device
When passing the device on to third parties, be sure to include the operating manual and
these instructions for use.
5 Disposal
Observe the applicable legal regulations when disposing of the device. The disposal is regulated
within the European Community by national regulations based on EU Directive 2002/96/EC on
waste electrical and electronic equipment (WEEE). According to these regulations, any devices
supplied after August 13, 2005 on a business-to-business basis, to which this product is
assigned, may no longer be disposed of in municipal or household waste. For more information,
please contact your supplier.
After installation, wait four hours before
switching the centrifuge on in order to avoid
damage to the compressor.
Check that the mains voltage and frequency
match the requirements on the device type
plate (see rear side of the device) and then
connect the device to the power supply.
1. Switch on the device at the mains power
switch.
2. Open the centrifuge lid using the open key.
3. Turn the rotor nut counterclockwise using the
rotor key supplied.
4. Remove rotor by lifting vertically.
5. Remove transport protection pad.
6. Place rotor vertically onto the motor shaft.
7. Turn the rotor nut with rotor key clockwise
until the rotor nut is tightened.