4.2 Methodensequenzen

Probenwechsler 760, Anhang

28


Methode 4

760 Sample Changer 760.0012
parameters
method 760_4
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 SAMPLE: - 1
3 SAMPLE: + 1
>sample sequence
1 MOVE 1 : sample
2 LIFT: 1 : work mm
3 STIR: * : ON s
4 CTL:Rm: START dos*
5 WAIT 10 s
6 CTL:Rm: START device1
7 SCN:Rm : End1
8 STIR: * : OFF s
9 LIFT: 1 : rinse mm
10 CTL:Rm: ****1*********
11 CTL:Rm: ****0*********
12 SCN:Rm : ******1*
13 LIFT: 1 : shift mm
14 WAIT 3 s
>final sequence
1 MOVE 1 : spec.1
2 LIFT: 1 : work mm
>changer settings
rack number 0
lift rate 1 25 mm/s
shift rate 20
shift direction: auto.
beaker test mode: single
on beaker error: MOVE
>stirring rates
stirrer 1 3
stirrer 2 3
>dosing unit def.
>manual stop
CTL Rmt: STOP device1
CTL RS232:
------------

¬

Report-Header mit Programmversion


¬

Methodenname

¬

Anzahl der abzuarbeitenden Proben (hier

ganzes Probenrack)
¬

Remote Schnittstelle initialisieren

¬

Initialisierung der

¬

Probenposition


¬

nächste Probe vor Turm fahren

¬

Lift mit Titrierkopf in Arbeitsposition fahren

¬

Rührer einschalten

¬

Dosieren starten

¬

Wartezeit: 10 Sek.

¬

Titrino starten

¬

Ende der Titration abwarten

¬

Rührer ausschalten

¬

Lift mit Titrierkopf in Spülposition fahren

¬

Externe Pumpe einschalten

¬

Remoteleitung zurücksetzen

¬

Pumpenbereitschaft abwarten

¬

Lift mit Titrierkopf in Drehposition fahren

¬

Wartezeit: 3 Sek. Abtropfen lassen


¬

Konditionierbecher vor Turm fahren

¬

Elektrode eintauchen

------ Einstellungen für Wechslerfunktionen --------




¬

bei fehlendem Probenbecher wird automatisch

der nächste angefahren
------ Rührgeschwindigkeiten ---------------------------


------ Einstellungen für Dosiergeräte (hier keine) --
------ Reaktion auf manuellen Abbruch --------------
¬

Gerät 1 stoppen