The following are suggested values:
Baud rate
9600
Data bits
8
Stop bit(s)
1
Parity
none
Handshake HWshort
Set the following parameters on the Titroprocessor if you have the PC con-
nected to the 796’s RS232 Interface 1.
ЪДДД Configuration: Interfaces ДДДДД¿
³ ³
³ RS232 Ifc.1 Ifc.2 ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ Baud rate 9600 9600 ³ <— Data transfer parameter
³ Data bit 8 8 ³ <— "
³ Stop bit 1 1 ³ <— "
³ Parity none none ³ <— "
³ Handshake HWshort HWshort ³ <— "
³ Receive mode off off ³ <— must be set off for file backup
³ ³
³ Balance none none ³
³ Printer none none ³
³ ³
³ Communication Int. 1 ³ <— RS232 interface
³ Commun. mode keyCTRL ³ <— Communication mode
³ ³
³ Internal printer on ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ Auto form feed off ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДД Quit ÄÙ