5.1.3
Reflected Light Measurement
Set the diffuser ring surrounding the retractable diffuser
head
to “Reflected light measurement”
for this
mode. The corresponding icon appears in the display.
6
Functions Menu
– Measuring/Operating Modes
Mode Selection: Ambient Light
– Flash Light
DIGISKY can be switched on using any key. All measuring and operating modes are
instantly available.
If none of the meter modes are activated for a period of time, the meter will be switched
into standby mode, and subsequently into sleep mode (setting option in Menu
–
Settings
– Display off or Power Off, see page 27).
The DIGISKY can only be switched on from the sleep mode by pressing the
measurement key (M).
Function Selection: Ambient light
– Flash light
Use the menu key ...
to select a function
Active: ambient light
reading mode
Active: flash mode
i
Only light reflected by the subject is measured. When
this metering method is used, results are always
dependent on reflection from the subject! This means,
for example, that intrinsically bright subjects will be
reproduced darker, and not as precisely exposed as with
the incident light method.
When employing this metering method, the use of a
suitable grey card (18% coefficient of diffuse reflection)
will help to deliver consistent results.
M
Measured f-stop increments (1 EV, 1/2 EV, 1/3 EV) can be adapted as required by the user in
Menu
– User (Cam 1, Cam 2, Cam 3, Cine) - EV Steps (see page 25).
i
The DIGISKY works with f-stop increments of 1/10 each. Measured correction values for the