INFICON SKY CDGxxxD (RS232C) Benutzerhandbuch
Seite 9

tira49d1 (2009-01) RS232C.kal
9
(abgeschlossen)
Parameter Name
Daten Typ
Beschreibung
Byte Nr. 2
Byte Nr. 3
Bemerkung
Extended Error
uint8_T / Ro
L-Byte
54
Bit 0
Atm.-Pressure out of range
L-Byte
Bit 1
Temperature out of range
Bit
2
Reserve
Bit
3
Reserve
Bit 4
Cal-Mode falsch
Bit 5
Pressure Underflow
Bit 6
Pressure Overflow
Bit
7
Zero-Adjust-Overflow
Extended Error
uint8_T / Ro
H-Byte
55
Bit 0
PT1000 fault
4)
H-Byte
Bit 1
Heaterblock overtemp.
4)
Bit
2
Reserve
Bit
3
Reserve
Bit
4
Reserve
Bit
5
Reserve
Bit
6
Reserve
Bit
7
Reserve
Pressure range
uint8_T / Ro
56
0
F.S.R. = E-3
(Exponent)
1
F.S.R.
=
E-2
2
F.S.R.
=
E-1
3
F.S.R. = E 0
4
F.S.R.
=
E+1
5
F.S.R.
=
E+2
6
F.S.R.
=
E+3
7
F.S.R.
=
E+4
Pressure range
uint8_T / Ro
57
0
Mantisse = 1.0
(Mantisse)
1
Mantisse = 1.1
2
Mantisse = 2.0
3
Mantisse = 2.5
4
Mantisse = 5.0
Gauge config
uint8_T / Ro
L-Byte
58
0
= Analogout 0 … 10.24 V
1
= Analogout 1 … 9 V
CDG type
uint8_T / Ro
L-Byte
59
0
= CDG025D
1
=
CDG045D
2
=
CDG100D
3
=
CDG160D
4
=
CDG200D
RW = Read / Write
Ro = Read only
1)
Werkseinstellung
2)
Umrechnung
→ Kapitel 1.1, Byte 6 "Lesebefehl"
3)
Umrechnung
→ Kapitel 1.1, Byte 4 und 5 "Druckeinheit"
4)
Nur CDG045D, CDG100D, CDG160D, CDG200D