1description, 1 general, 2 technical data – INFICON VSA100A differential to ambient Benutzerhandbuch

Seite 11

Advertising
background image

1

Description

1.1

General

Unpack the Vacuum Switch VSA100A immediately after
delivery, even if it is to be put into operation at a later
date.

Examine the transport box for any external damage.
Then completely remove the packaging materials.

Note
The transport box and packaging materials must be kept
in the event of complaints about damage.

Check that the Vacuum Switch VSA100A is complete
and carefully examine it visually to ensure that no
damage has occurred during transport.

If any damage is discovered, report it immediately to the
forwarding agent and insurer. If the damaged part has to
be replaced, please get in touch with our Customer
Support.

1.1.1 Intended Use

The Vacuum Switch VSA100A is used as a pressure
switch in vacuum systems. It is used e.g. to automatical-
ly interrupt the gas supply when venting vacuum sy-
stems with a purge gas at a pressure of 3 mbar below at-
mospheric pressure.

1.2

Technical Data

Switching pressure

6 mbar

(below atmospheric pressure)

Return switching pressure

3 mbar

(below atmospheric pressure)

Switching inaccuracy

± 2 mbar

Max. operating pressure (abs)

2000 mbar

Operating temperature

0 to 85

°

C

Helium permeation

< 1·10

-6

mbar·l·s

-1

Switching contact

changeover contact

Switching capacity

100 mA / 24 VAC

30 mA / 24 VDC

11

11

tina29d2-b (2004-10) VSA100A.ga

tina29d2-b (2004-10) VSA100A.ga

1

Description

1.1

General

Unpack the Vacuum Switch VSA100A immediately after
delivery, even if it is to be put into operation at a later
date.

Examine the transport box for any external damage.
Then completely remove the packaging materials.

Note
The transport box and packaging materials must be kept
in the event of complaints about damage.

Check that the Vacuum Switch VSA100A is complete
and carefully examine it visually to ensure that no
damage has occurred during transport.

If any damage is discovered, report it immediately to the
forwarding agent and insurer. If the damaged part has to
be replaced, please get in touch with our Customer
Support.

1.1.1 Intended Use

The Vacuum Switch VSA100A is used as a pressure
switch in vacuum systems. It is used e.g. to automatical-
ly interrupt the gas supply when venting vacuum sy-
stems with a purge gas at a pressure of 3 mbar below at-
mospheric pressure.

1.2

Technical Data

Switching pressure

6 mbar

(below atmospheric pressure)

Return switching pressure

3 mbar

(below atmospheric pressure)

Switching inaccuracy

± 2 mbar

Max. operating pressure (abs)

2000 mbar

Operating temperature

0 to 85

°

C

Helium permeation

< 1·10

-6

mbar·l·s

-1

Switching contact

changeover contact

Switching capacity

100 mA / 24 VAC

30 mA / 24 VDC