tina29d2-b (2004-10) VSA100A.ga

2

Start up

Note
When leak detecting with a helium leak detector,
spraying the pressure switch with helium may result in a
helium proof. This is in general not due to a leak of the
VSA100A, but is permeation of helium through the
EPDM diaphragm.

2.1

Vacuum Connection

The VSA100A is connected to the vacuum system via a
DN 16 KF small flange.

The VSA100A should preferably be mounted standing
on its flange. Inclined mounting is also permissible,
including the horizontal position. Downward installation
is not permissible because in such a case condensate
may collect in the pressure switch. This affects the
measurements and may damage the pressure switch.

Dimensional diagram of VSA100A, see Fig. 1.

2.2

Electrical Connection

The contact assignment is given in Fig. 2.

The VSA100A is supplied with a connected 3 m long

tina29d2-b (2004-10) VSA100A.ga

13

13

72

50

DN16 KF

Ø

48.5

Fig. 1 Dimensional diagram VSA100A

Atmospheric pressure
(rest position)

Fig. 2 Contact assignment of microswitch

1 (black)

1 (black)

3 (brown)

2 (grey)

2 (grey)

3 (brown)

Low pressure
(operating position)

2

Start up

Note
When leak detecting with a helium leak detector,
spraying the pressure switch with helium may result in a
helium proof. This is in general not due to a leak of the
VSA100A, but is permeation of helium through the
EPDM diaphragm.

2.1

Vacuum Connection

The VSA100A is connected to the vacuum system via a
DN 16 KF small flange.

The VSA100A should preferably be mounted standing
on its flange. Inclined mounting is also permissible,
including the horizontal position. Downward installation
is not permissible because in such a case condensate
may collect in the pressure switch. This affects the
measurements and may damage the pressure switch.

Dimensional diagram of VSA100A, see Fig. 1.

2.2

Electrical Connection

The contact assignment is given in Fig. 2.

The VSA100A is supplied with a connected 3 m long

72

50

DN16 KF

Ø

48.5

Fig. 1 Dimensional diagram VSA100A

Atmospheric pressure
(rest position)

Fig. 2 Contact assignment of microswitch

1 (black)

1 (black)

3 (brown)

2 (grey)

2 (grey)

3 (brown)

Low pressure
(operating position)