Seite 83: 13 kommunikations-schemas für automation, Measurement

2.13
Kommunikations-Schemas für Automation
VA Computrace 797 / Hardware-Gebrauchsanweisung 8.797.8001DE
75
2.13.5
Kommunikationsschema "RC Record response curve"
797 VA Computrace
Dosinos 1..3
838 Advanced Sample
Processor
731 Relay Box
Set line: Start 838 (TTL 3)
Ready (TTL 7)
Set line: Siphoning (TTL 5)
Prep Dosino 1,2,3 ?
Prep Dosino 1, 2, 3
YES
Set line: siphoning (TTL 5)
Wait siphoning time
Set line: Rinsing (TTL 6)
Wait rinsing time
No. of Prep
cycles
Is it the
first sample of the
sequence ?
NO
Set line: Siphoning (TTL 5)
Wait siphoning time
NO
YES
Set line: Sample transfer (TTL 4)
Ready (TTL 7)
Predose
Measurement
Set line: Siphoning (TTL 5)
Wait siphoning time
Set line: Rinse needle (TTL 4)
Ready (TTL 7)
Set line: Siphoning (TTL 5)
Wait siphoning time
Set line: Rinsing (TTL 6)
Wait rinsing time
No. of rinsing cycles
Export/Print results
Set line: End of sample (TTL 1)
More samples ?
NO
YES
Set line: End of sample table (TTL 2)
RC Record response curve
<Use for predose> active ?
YES
NO
Scan line : 838 ready
(TTL 7)
Scan line : 838 ready
(TTL 7)
Scan line : 838 ready
(TTL 7)
No. of rinsing cycles
Wait siphoning time
Set line: Siphoning (TTL 5)
Wait siphoning time
Note
: Use <Use
for predose> to
dose Electrolyte
via Dosino 1
Ti
m
e
line
End
Abb. 41:
Kommunikations-Schema für "RC Record response curve"