Rc sample with response curve, 2 installation, Measurement – Metrohm 797 VA Computrace Benutzerhandbuch

Seite 84: Measure q(0), Measurement end

Advertising
background image

2 Installation

VA Computrace 797 / Hardware-Gebrauchsanweisung 8.797.8001DE

76

2.13.6

Kommunikationsschema "RC Sample with response curve"

797 VA Computrace

Dosinos 1..3

838 Advanced Sample

Processor

731 Relay Box

Set line: Start 838 (TTL 3)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Prep Dosino 1,2,3 ?

Prep Dosino 1, 2, 3

YES

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of Prep
cycles

Is it the

first sample of the

sequence ?

NO

Set line: Siphoning (TTL 5)

Wait siphoning time

NO

YES

Set line: Transfer sample (TTL 4)

Ready (TTL 7)

Predose

Measurement

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinse Capillary (TTL 4)

Ready (TTL 7)

Set line: Siphoning (TTL 5)

Wait siphoning time

Set line: Rinsing (TTL 6)

Wait rinsing time

No. of rinsing cycles

Export/Print results

Set line: End of sample (TTL 1)

More samples ?

NO

YES

Set line: End of sample table (TTL 2)

RC Sample with response curve

<Use for predose>

active ?

YES

NO

No. of rinsing cycles

Wait siphoning time

Set line: Siphoning (TTL 5)

Wait siphoning time

Measure Q(0)

Dose

<Use after

sample transfer>

active ?

YES

NO

Note: Use <Use

for predose> to

dose Electrolyte

via Dosino 1

T

ime

lin

e

Measurement end

Set line: Siphoning (TTL 5)

Wait siphoning time

Scan line: Waiting for

START signal

Scan line: Waiting for

START signal

Scan line: Waiting for

START signal

Remove Elect-

rolyte before adding the pro-

duction bath ?

NO

YES

Abb. 42:

Kommunikations-Schema für "RC Sample with response curve"

Advertising