3.2 Remote control commands
751 GPD Titrino
153
3.2.2.51. Mode.Parameter.Presel.RateDisp ON,
OFF
Parameters for STAT, DOC: Display of the current rate during the determina-
tion.
3.2.2.52. Mode.Parameter.DosPara.Type volume, time, rate
Mode.Parameter.DosPara.Volume.Volume 0.000...10...99999.99
Mode.Parameter.DosPara.Volume.DisType time, rate
Mode.Parameter.DosPara.Volume.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Volume.Time 1...100...999 999
Mode.Parameter.DosPara.Time.Time 1...100...999 999
Mode.Parameter.DosPara.Time.DisType volume, rate
Mode.Parameter.DosPara.Time.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Time.Volume 0.000...10...99999.99
Mode.Parameter.DosPara.Rate.Rate 0.001...150,
max.
Mode.Parameter.DosPara.Rate.StopType volume, time
Mode.Parameter.DosPara.Rate.Time 1...100...999 999
Mode.Parameter.DosPara.Rate.Volume 0.000...10...99999.99
Parameters for DOS: In the equation Rate = Volume/Time, two parameters
are preset, the third is calculated. Depending on the entry of the type, the fol-
lowing information is required: .Volume, .Time or .Rate.
Entries: .Volume in ml, .Time in s, .Rate in ml/min.
3.2.2.53. Mode.Parameter.DosPara.Pause
0...999 999
Parameters for DOS: Pause time in s.
3.2.2.54. Mode.Parameter.DosPara.TDelta 1...10...999 999
Parameters for DOS, DOC: Time interval for the entry of a measured point in
the list of measured points.
3.2.2.55. Mode.Parameter.DosPara.Temp -170.0...25.0...500.0
Parameters for DOS, DOC: Temperature in °C. The temperature is measured
continuously and updated if a T sensor is attached.
The temperature is used for temperature compensation in pH measurements.
3.2.2.56. Mode.Parameter.Monitoring.MeasVal.MeasModepH,U,Ipol,Upol,OFF
Mode.Parameter.Monitoring.MeasVal.MeasInput
1, 2, diff.
Mode.Parameter.Monitoring.MeasVal.PCurrent 0...1...±127
Mode.Parameter.Monitoring.MeasVal.PVoltage 0...400...±1270
Mode.Parameter.Monitoring.MeasVal.PolElectrTest ON, OFF
Mode.Parameter.Monitoring.MeasVal.pH.LowLim
0...±20.00
Mode.Parameter.Monitoring.MeasVal.pH.UpLim 0...±20.00
Mode.Parameter.Monitoring.MeasVal.U.LowLim
0...±2000
Mode.Parameter.Monitoring.MeasVal.U.UpLim 0...±2000
Mode.Parameter.Monitoring.MeasVal.Ipol.LowLim
0...±2000
Mode.Parameter.Monitoring.MeasVal.Ipol.UpLim 0...±2000