3.2 Remote control commands

751 GPD Titrino

153

3.2.2.51. Mode.Parameter.Presel.RateDisp ON,

OFF

Parameters for STAT, DOC: Display of the current rate during the determina-

tion.

3.2.2.52. Mode.Parameter.DosPara.Type volume, time, rate

Mode.Parameter.DosPara.Volume.Volume 0.000...10...99999.99

Mode.Parameter.DosPara.Volume.DisType time, rate

Mode.Parameter.DosPara.Volume.Rate 0.001...150,

max.

Mode.Parameter.DosPara.Volume.Time 1...100...999 999

Mode.Parameter.DosPara.Time.Time 1...100...999 999

Mode.Parameter.DosPara.Time.DisType volume, rate

Mode.Parameter.DosPara.Time.Rate 0.001...150,

max.

Mode.Parameter.DosPara.Time.Volume 0.000...10...99999.99

Mode.Parameter.DosPara.Rate.Rate 0.001...150,

max.

Mode.Parameter.DosPara.Rate.StopType volume, time

Mode.Parameter.DosPara.Rate.Time 1...100...999 999

Mode.Parameter.DosPara.Rate.Volume 0.000...10...99999.99

Parameters for DOS: In the equation Rate = Volume/Time, two parameters

are preset, the third is calculated. Depending on the entry of the type, the fol-

lowing information is required: .Volume, .Time or .Rate.

Entries: .Volume in ml, .Time in s, .Rate in ml/min.

3.2.2.53. Mode.Parameter.DosPara.Pause

0...999 999

Parameters for DOS: Pause time in s.

3.2.2.54. Mode.Parameter.DosPara.TDelta 1...10...999 999

Parameters for DOS, DOC: Time interval for the entry of a measured point in

the list of measured points.

3.2.2.55. Mode.Parameter.DosPara.Temp -170.0...25.0...500.0

Parameters for DOS, DOC: Temperature in °C. The temperature is measured

continuously and updated if a T sensor is attached.

The temperature is used for temperature compensation in pH measurements.

3.2.2.56. Mode.Parameter.Monitoring.MeasVal.MeasModepH,U,Ipol,Upol,OFF

Mode.Parameter.Monitoring.MeasVal.MeasInput

1, 2, diff.

Mode.Parameter.Monitoring.MeasVal.PCurrent 0...1...±127

Mode.Parameter.Monitoring.MeasVal.PVoltage 0...400...±1270

Mode.Parameter.Monitoring.MeasVal.PolElectrTest ON, OFF

Mode.Parameter.Monitoring.MeasVal.pH.LowLim

0...±20.00

Mode.Parameter.Monitoring.MeasVal.pH.UpLim 0...±20.00

Mode.Parameter.Monitoring.MeasVal.U.LowLim

0...±2000

Mode.Parameter.Monitoring.MeasVal.U.UpLim 0...±2000

Mode.Parameter.Monitoring.MeasVal.Ipol.LowLim

0...±2000

Mode.Parameter.Monitoring.MeasVal.Ipol.UpLim 0...±2000