3.2 Remote control commands
751 GPD Titrino
175
The beginning of the message is marked by a space (ASCII 32).
3.2.2.163. Setup.Lock.Keyboard ON, OFF
Setup.Lock.Config ON, OFF
Setup.Lock.Parameter ON, OFF
Setup.Lock.SmplData ON, OFF
Setup.Lock.UserMeth.Recall ON, OFF
Setup.Lock.UserMeth.Store ON, OFF
Setup.Lock.UserMeth.Delete ON, OFF
Setup.Lock.Display ON, OFF
ON means disable the corresponding function:
.Keyboard: Disable all keys of the Titrinos
.Config:
Disable the <configuration> key
.Parameter: Disable the <parameter> key
.SmplData: Disable the <smpl data> key
.UserMeth.Recall: Disable "recall" in <user meth> key
.UserMeth.Store: Disable "store" in <user meth> key
.UserMeth.Delete: Disable "delete" in <user meth> key
.Display: Disable the display, i.e. it will not be written to by the device pro-
gram of the Titrino and can be operated from the computer.
3.2.2.164. Setup.Mode.StartWait ON, OFF
Setup.Mode.FinWait ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method or submethod in TIP
(holding point after AutoInfo !".T.GC").
.FinWait: Holding point at the end a method or submethod in TIP (holding
point after AutoInfo !".T.F").
3.2.2.165. Setup.SendMeas.SendStatus ON, OFF
Setup.SendMeas.Interval 0.08...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
3.2.2.167 and 3.2.2.168) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.167 and 3.2.2.168.
The inputted value is rounded off to a multiple of 0.08. The small-
est possible time interval depends on the number of measured
values which have to be sent, on the baud rate, on the load on the
interface and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.
The automatic transmission is switched on/off with 'SendStatus'.
3.2.2.166. Setup.SendMeas.Select Assembly,
Titrator
Selection of the unit of which the measured values should be sent (3.2.2.167
or 3.2.2.168).