WIKA Operating Instructions Temperature Transmitter T32.10/11/30
3. Intended use
Electromagnetic compatibility per EN 61 326 and NAMUR recom-
mendation NE21
Signalling at the analogue output per NAMUR recommendation
NE3
Sensor burnout signalling per NAMUR recommendation NE89
Before initial operation check the suitability for the intended
application.
3.1 Functional description
The temperature transmitter serves for the conversion of a
resistance value or a voltage value into a proportional current signal
( ... 20 mA). The analogue signal is transmitted to a series-
connected logic unit, e.g. an SPS or an alarm contact, where
it is monitored for upper deviations from a maximum value or
lower deviations from a minimum value. For failure monitoring the
logic unit has to be capable of recognising HI alarms (adjustable
from 21 ... 22. mA) as well as LO alarms (3.6 mA). The electrical
components of the transmitter are arranged in a plastic case and
completely encapsulated.
3.2 Ambient conditions
T32.1*.***
T32.3*.***
Ambient/storage
temperature
Standard range
Extended range
-0 ... +8 °C
-0 ... +8 °C
-20 xxx +70 °C
---
Climate class
Cx (-0 ... +8 °C, %
to 9 % rel. air humidity)
DIN EN 60 6-1
Bx (-20 ... +70 °C, %
to 9 % rel. air humidity)
DIN EN 60 6-1
Max. permissible
humidity
100 % relative humidity
(unlimited with insulated
sensor connection
wires), moisture conden-
sation permissible
DIN IEC 68-2-30 Var. 2
99 % relative humidity
DIN IEC 68-2-30 Var. 2
Vibration
10 ... 2000 Hz g DIN IEC 68-2-6