INFICON LDS3000 Modular Leak Detector Benutzerhandbuch
Seite 25

Betrieb
25
9
1
Pressure p1 log. in Pa
1 V/decade, 2,5 ... 8,5 V,
2,5 V = 1E-3 mbar
p1 1 V/dec.
don't care don't care
9
2
Pressure p2 log. in Pa
1 V/decade, 2,5 ... 8,5 V,
2,5 V = 1E-3 mbar
p2 1 V/dec.
don't care don't care
10
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-3 mbar·l/s
Leak rate log.
2 V/dec.
1E+2 mbar·l/s
10
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-3 mbar·l/s
Leak rate log.
special_1 1E+1 mbar·l/s
11
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-4 mbar·l/s
Leak rate log.
2 V/dec.
1E+1 mbar·l/s
11
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0V = 1E-4 mbar·l/s
Leak rate log.
special_1 1E+0 mbar·l/s
12
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-5 mbar·l/s
Leak rate log.
2 V/dec.
1E0 mbar·l/s
12
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-5 mbar·l/s
Leak rate log.
special_1 1E-1 mbar·l/s
13
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-6 mbar·l/s
Leak rate log.
2 V/dec.
1E-1 mbar·l/s
13
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-6 mbar·l/s
Leak rate log.
special_1 1E-2 mbar·l/s
14
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-7 mbar·l/s
Leak rate log.
2 V/dec.
1E-2 mbar·l/s
14
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-7 mbar·l/s
Leak rate log.
special_1 1E-3 mbar·l/s
15
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-8 mbar·l/s
Leak rate log.
2 V/dec.
1E-3 mbar·l/s
15
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-8 mbar·l/s
Leak rate log.
special_1 1E-4 mbar·l/s
16
1
Leak rate log. in mbar·l/s
0 ... 8 V, 2 V/decade,
0 V = 1E-9 mbar·l/s
Leak rate log.
2 V/dec.
1E-4 mbar·l/s
16
2
Leak rate log. in mbar·l/s
0 ... 10 V, 3 V/decade,
0 V = 1E-9 mbar·l/s
Leak rate log.
special_1 1E-5 mbar·l/s
LDS2010
setting
menu item
22
Analog
output
channel
Function LDS2010
Function LDS3000
Scale for
leakrate
upper limit
(10 V = ...)