Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 225

Report templates
199
Calibration curve field
Calibration curve fields are used for showing calibration or standard addition
curves in the report.
Insert
In order to insert a calibration curve field in a report template the corre-
sponding symbol must be selected on the Module bar and then placed on
the report template by creating a field with the left-hand mouse key.
Properties
X pos.
0.0000 ... (max. page width) mm
x-position within the permitted section.
Y pos.
0.0000 ... (max. page height) mm
y-position within the permitted section.
Width
0.0000 ... (max. page width) mm
Width of calibration curve field.
Height
0.0000 ... (max. page height) mm
Height of calibration curve field.
Command name
[ not defined ], 50 characters
Entry of the name of the command for which the calibration curve is to be
produced. With
not defined
the first calibration curve present in the determi-
nation with the selected
Command type
will be produced as default.