Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 523

Commands
497
•
Standard addition
Parameters for standard addition.
•
Measuring parameters
Parameters for the measurement.
Standard addition curve
The standard addition curve with its evaluation data for a determination is shown
in the program part Database in the dialog window
Calibration curve
which is
opened with the menu item
Determinations, Show calibration curve...
or the symbol
.
General/Hardware
The general parameters for the control device, the dosing device, the sensor and
the stirrer are defined on this tab.
Command name
25 characters, [ STDADD dos # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.