Commands
345
Stirring rate
-15 ... [ 8 ] ... 15
(only Titrando)
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically
[ on ], off
(only Titrando)
If this option is enabled, the stirrer will be switched off automatically when
the command is finished.
Switch on/off automatically
[ on ], off
only 751, 785, 798, 799)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command.
Start conditions
The start conditions are processed in the listed sequence before the titration is
started.
Command name
25 characters, [ DET U # ]
Name of the command.
Initial measured value
The initial measured value is determined before the other start conditions are
processed. The measured value acceptance depends on the three following pa-
rameters:
Signal drift
0.1 ... 999.0 mV/min, [ off ]
(only Titrando)
The measured value is only accepted if the drift undercuts the entered
value.
off
means that the measured value is accepted after the maximum
waiting time.