Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 457

Commands
431
Titration parameters
Parameters defining the run of the titration.
Command name
25 characters, [ STAT pH # ]
Name of the command.
Titration direction
Selection of the titration direction.
+
Positive measured value alteration.
[ - ]
Negative measured value alteration.
auto
The titration direction is determined automatically from the start measured
value and the set control point.
Temperature
-20.0 ... [ 25.0 ] ... 150.0 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500.0 °C
(Titrino)
Manually entered titration temperature. If a temperature sensor is con-
nected and the
Temperature measurement
in the tab
General/Hardware
under
Sensor
is set to
automatic
or
continuous
then the temperature will be meas-
ured continuously.
Measuring point recording
Time interval measuring point
0.1 ... [ 2.0 ] ... 999999 s
(Titrando)
1 ... [ 2 ] ... 999999 s
(Titrino)
Time interval for entering a measuring point in the list of measuring points.
Start time
[ 0 ] ... 999999 s
The measured values are not entered in the list of measuring points until
this time has elapsed since the start of the titration.
Start measured value pH
-20.000 ... 20.000, [ off ]
(Titrando)
-20.00 ... 20.00, [ off ]
(Titrino)
The measured values are not entered in the list of measuring points until
this value has been reached.
Start rate
0.01 ... 166.00 mL/min, [ off ]
(Titrando)
0.01 ... 150.00 mL/min, [ off ]
(Titrino)