Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 728

702
Configuration
Edit rack properties (774)
Rack name
Shows the name of the attached rack. If no rack is in position then "
-----
" will
be shown.
Rack code
Shows the rack code of the attached rack. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the
Sample Processor in order to recognize the rack. If no rack is in position
then the display will be empty.
Number of positions
Shows the number of positions on the attached rack. If no rack is in posi-
tion then the display will be empty.
The parameters for the attached rack are defined on the following 2 tabs:
•
Lift positions
Defines the work, rinse, shift and special positions for Tower 1.
•
Special beaker
Settings for the special beaker on the rack in position.
Lift positions
Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions ex-
cept those that are defined as Special beaker.
Work position
0 ... [ 83 ] ... 100 mm
Work position for Lift 1. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.
Rinse position
[ 0 ] ... 100 mm
Rinse position for Lift 1. This lift position is used for rinsing the electrodes.