Commands
529
Monitoring
Command name
25 characters, [ DOS U # ]
Name of the command.
Monitoring Measured value
on, [ off ]
If this option is switched on then the measured value will be monitored and
any infringements will be entered in the list of measuring points.
Lower limit
[ -2000.0 ] ... 2000.0 mV
(Titrando)
[ -2000 ] ... 2000 mV
(Titrino)
Lower limit of the measured value. When the measured value falls below
this value, the action
Measured value lower limit exceeded
is triggered.
Lower hysteresis
0.0 ... [ 2.0 ] ... 2000.0 mV
(only Titrando)
The lower hysteresis represents a tolerance range for the lower limit of the
measured value. When the measured value exceeds the lower limit plus this
hysteresis value, the action
Measured value lower limit ok
is triggered.
Upper limit
-2000.0 ... [ 2000.0 ] mV
(Titrando)
-2000 ... [ 2000 ] mV
(Titrino)
Upper limit of the measured value. When the measured value exceeds this
value, the action
Measured value upper limit exceeded
is triggered.