Configuration
life of the sensor has expired will be saved automatically with the determi-
nation.
Stop determination
The running determination will be terminated automatically. The following
message must be confirmed with
[OK]
.
Calibration data
Note
This tab only appears for pH and ISE electrodes.
Slope
pH electrode:
-999.9 ... 999.9 %, [ 100.0 ]
ISE electrode:
-999.9 ... 999.9 mV, [ 59.2/valency ]
Slope of the pH or ISE electrode. This value can be determined automati-
cally with a calibration from the linearized calibration curve or entered
manually. The slope of a sensor that is used by the commands
DET
,
MET
,
SET
,
KFT
,
MEAS
and
CAL LOOP
is available as a variable
'Command name.SLO'
for calculations.
pH(0)
-20.000 ... [ 7.000 ] ... 20.000
The electrode zero point
pH(0)
is only shown for pH electrodes. After the
slope the
pH(0)
is the second characteristic quantity of the calibration func-
tion U = f(pH).
pH(0)
is the x-axis intercept of the calibration curve, i.e. it
corresponds to the pH value at 0 mV. The electrode zero point
pH(0)
of a
pH electrode that is used by the commands
DET
,
MET
,
SET
,
KFT
,
MEAS
and
CAL LOOP
is available as a variable
'Command name.ENP'
for calculations.
E(0)
-2000.0 ... [ 0.0 ] ... 2000.0 mV
The electrode zero point
E(0)
is only shown for ISE electrodes.
E(0)
is the
second characteristic quantity of the calibration function U = f(log c).
E(0)
is the y-axis intercept of the calibration curve and corresponds to the poten-
tial at log c = 0. The electrode zero point
E(0)
of an ISE electrode that is
used by the commands
DET
,
MET
,
SET
,
KFT
,
MEAS
and
CAL LOOP
is available
as a variable
'Command name.ENP'
for calculations.