Method
Control parameters
Command name
25 characters, [ KFC # ]
Name of the command.
Endpoint
EP at
-2000.0 ... [ 50.0 ] ... 2000.0 mV
Measured value for the end point.
Control
Dynamics
1.0 ... [ 70 ] ... 2000 mV, off
The dynamics is entered as distance to the end point. Within the dynamics
iodine will be produced with
Min. rate
. The bigger the control range the
slower the titration. Outside the dynamics iodine will be produced continu-
ously with
Max. rate
.
The standard control parameters are optimal for most applications and
should not be altered. If you nevertheless need to alter the control parame-
ters for special reagents and/or samples take care that the polarization cur-
rent of the indicator electrode (
I(pol)
), the end point and the dynamics are
linked to each other.