10 result commands, Calc – Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 578

552
Method
Gas
Gas type
[ air ], nitrogen, other gas
Gas type of the carrier gas.
Gas flow factor
0.001 ... [ 1.000 ] ... 9.999
Factor for the correct determination of the gas flow. This parameter is only
visible for
Gas type
=
other gas
. The values for frequently used carrier gases
are listed in the Instructions for Use for the 774 Oven Sample Processor.
Gas flow
Inlet
Selection of the inlet of the carrier gas:
[ pump ]
Use of ambient air as carrier gas.
valve
Use of gas out of a compressed gas cylinder as carrier gas.
Switch on
[ on ], off
Switch on the gas flow.
Switch off
on, [ off ]
Switch off the gas flow.
5.6.10 Result
commands
Commands for the calculation, storage and output of results.
The following four result commands can be selected:
•
CALC
Calculation of intermediate and end results, titer values and common vari-
ables.
•
DATABASE
Storage of the determination data in databases.
•
EXPORT
Export of determination data.
•
REPORT
Output of a report defined by a report template.
CALC
Command for the calculation of intermediate and end results and the as-
signment of titer values and common variables.
Appearance
The command has the following appearance: