488
Method
General/Hardware
The general parameters for the control device and the sensor are defined on this
tab.
Command name
25 characters, [ MEAS Cond # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a 712 Conductometer from the device table. Additionally,
not
defined
can be selected. In this case, the device name must be assigned by
the user at the start of the method.
Device type
712 Conductometer
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Sensor
Sensor
Sensor name, [ Conductivity sensor ]
Selection of a sensor of type
Conductivity sensor
from the sensors config-
ured in the sensor table.