Subwindow Devices
673
Rack
On the tab
Rack
the rack-specific data of the attached rack are shown.
Rack name
Shows the name of the attached rack. If no rack is in position then "
-----
" will
be shown.
Rack code
Shows the rack code of the attached rack. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the
Sample Processor in order to recognize the rack. If no rack is in position
then the display will be empty.
Number of positions
Shows the number of positions on the rack. If no rack is in position then the
display will be empty.
Shift rate
3 ... [ 20 ] °/s
Sample rack speed of rotation for manual operation.
Opens the dialog window
Rack data
(see Configuration - Subwindow Rack
data - Rack properties) or
Rackdata (774)
in which the data of the attached
rack is shown and can be edited.
Initializes the attached rack. This resets the rack, lift and swing head, reads
off the rack code and transfers the corresponding rack data to the Sample
Processor.