Commands
429
Pause 1
Pause 1
[ 0 ] ... 999999 s
(only Titrando)
Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.
Start volume
Start volume
[ 0.00000 ] ... 9999.99 mL
(Titrando)
[ 0.00 ] ... 999.99 mL
(Titrino)
Volume to be added before the start of the titration with the below entered
speed.
Dosing rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando)
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.
Pause 2
Pause 2
[ 0 ] ... 999999 s
Waiting time, e.g. until the electrode has settled down after the start or a re-
action time after the addition of a start volume. The pause takes place at
the end of all start conditions.
Control parameters
Command name
25 characters, [ STAT pH # ]
Name of the command.
Control point
Control point at pH
-20.000 ... 20.000, [ off ]
(Titrando),
-20.00 ... 20.00, [ off ]
(Titrino)
Definition of the pH value to be controlled.