Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 486

460
Method
Evaluations
The following methods for evaluating measuring curves are defined on this tab.
Command name
25 characters, [ MEAS pH # ]
Name of the command.
Fix end point evaluation
on, [ off ]
To a fixed quantity (
Measured value
or
Time
) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at pH
-20.000 ... 20.000, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.
Fix EP# at
0.0 ... 999'999.9 s, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Time
.
The fixed end points Fix EP4...9 are displayed additionally to Fix EP1...3 by
pressing this button.
Only the fixed end points Fix EP1...3 are displayed by pressing this button.