Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 712

686
Configuration
Opens the dialog window
Load new program version
.
Old version
Shows the old device program version.
New version
Shows the new device program version to be loaded.
Loads new program version.
Note
Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.
Tower #
Tower parameters
Max. stroke path
0 ... [ 235 ] mm
Entry of the lowest permitted lift position. A lift height of
0 mm
corresponds
to the "Rest position", i.e. the lift is moved right to the top.
Min. beaker radius
1.0 ... 100.0 mm, [ off ]
Definition of the minimum radius which the beakers used on the rack must
have. If the lift is moved to the work position the beaker radii defined in
Rack table for general Sample positions and Special beaker will be com-
pared with the
Min. beaker radius
. If this minimum beaker radius is undercut
then the sequence will be stopped and an error message will appear. With
off
no check will be made.
Lift rate
3 ... [ 25 ] mm/s
Lift speed for manual operation.