General program functions
Note
The filling rate should be reduced for viscous liquids.
Fill automatically
This parameter is only visible for Titrandos, Dosing Interface and the USB
Sample Processor.
[ on ], off
If this option is switched on then the buret will be refilled automatically after
dosing. During the filling procedure the volume display is reset to
0.0000
mL
. If this option is switched off then the added volume will be displayed
cumulatively.
Start adding the fixed volume for the selected dosing device. The added
volume is shown live.
Note
Parameters that are altered after dosing has started only apply to the next dosing
procedure.
Start filling the buret for the selected dosing device. This button is only pre-
sent when
Fill automatically
is switched off. During the filling procedure the
volume display is reset to
0.0000 mL
.
Stop adding the fixed volume for the selected dosing device. Once dosing
has been stopped it can no longer be restarted.