Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 419

Commands
393
Control parameters
The Control parameters can be set for each endpoint separately.
Command name
25 characters, [ SET U # ]
Name of the command.
Endpoint 1
EP1 at
-2000.0 ... 2000.0 mV, [ off ]
(Titrando)
-2000 ... 2000, [ off ]
(Titrino)
Measured value for the first endpoint.
Titration rate
Titration rate
Three predefined sets of parameters can be selected for the titration rate:
slow
,
optimal
and
fast
. For these sets the parameters for the
Control
are not
displayed. In order to edit these parameters the value
user
must be se-
lected.
slow
For titrations in which the finest details are also to be visible. However, this
can also increase noise and lead to the production of unwanted EPs.
[ optimal ]
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical rapid titrations.
user
Editing the individual titration parameters which affect the titration rate.