Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 378

352
Method
Lower limit
[ -2000.0 ] ... 2000.0 mV
(for
Window
=
Measured value
)
[ 0.00000 ] ... 9999.99 mL
(for
Window
=
Volume
)
[ 0 ] ... 999999 s
(for
Window
=
Time
)
Value for the lower limit of the window.
Upper limit
-2000.0 ... [ 2000.0 ] mV
(for
Window
=
Measured value
)
0.00000 ... [ 9999.99 ] mL
(for
Window
=
Volume
)
0 ... [ 999999 ] s
(for
Window
=
Time)
Value for the upper limit of the window.
Gran evaluation
on, [ off ]
(only Titrando)
With the Gran evaluation titration curves are evaluated according to the
Gran plot procedure (see Titration commands - Evaluation - Gran evalua-
tion).
Initial volume
0.01 ... [ 50.00 ] ... 9999.99 mL
Volume which must be present in the measuring vessel before the com-
mand is started.
Lower limit
[ -2000.0 ] ... 2000.0 mV
Value for the lower limit of the Gran evaluation range.
Upper limit
-2000.0 ... [ 2000.0 ] mV
Value for the upper limit of the Gran evaluation range.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
DET Ipol
Command for Dynamic Equivalence Point Titrations with voltametric meas-
urement with (selectable polarization current).
Devices
This command can be executed with the following devices which have the DET
mode:
Titrino: 716, 721, 736, 751, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 855, 857
Appearance
The command has the following appearance:
Parameters
The parameters for the command
DET Ipol
are configured in the following 7 tabs:
•
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.