Prep – Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 561

Commands
535
Volume
0.00000 ... [ 10.00000 ] ... 99'999.9 mL (Dosing)
0.00000 ... [ 10.00000 ] ... 50.0 mL (Aspirate)
Enter the volume to be processed. This parameter is only visible for the
Functions
Dosing
and
Aspirate
.
Rate
[ maximum ]
0.01 ... 166.00 mL/min
Speed at which the commands
Dosing
,
Fill
,
Aspirate
,
Eject
and
Exchange po-
sition
are carried out. The maximum rate depends on the cylinder volume of
the Dosing unit used. Aspirating and ejecting the sample should be done
with a rate < 10 mL/min.
PREP
Command for rinsing cylinder and tubing of an exchange or dosing unit.
Devices
This command can be executed with the following devices:
Titrino: 736, 751, 758, 799
Titrando: 808, 809, 835, 836, 841, 842, 855, 857
Dosing Interface: 846
Sample Processor: 730, 774, 778, 789, 814, 815, 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command
PREP
are configured in the following dialog win-
dow:
•
PREP