405
Minimum evaluation
on, [ off ]
For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation).
Threshold value
0.5 ... [ 5.0 ] ... 10.0 µA/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.
Maximum evaluation
on, [ off ]
For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation.
Threshold value
0.5 ... [ 5.0 ] ... 10.0 µA/mL
The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.