Commands
489
Measuring parameters
The measuring parameters are defined on this tab.
Command name
25 characters, [ MEAS Cond # ]
Name of the command.
Measurement
Measurement frequency
[ auto ], 300 Hz, 2.4 kHz
The more suitable frequency is chosen with the setting
auto
.
Measuring time
0 ... [ 120 ] ... 999'999 s
Maximum period of time for measurement.
Measuring interval
1 ... [ 2 ] ... 999'999.0 s
Time interval for entering a measuring point in the list of measuring points.
Stop measured value
0.0000 ... 2000.0 mS/cm, [ off ]
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Temperature
Reference temperature
-170.0 ... [ 20.0 ] ... 500.0 °C
The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).
Measurement temperature
-170.0 ... [ 20.0 ] ... 500.0 °C
Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.