Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 430

404
Method
Additional evaluations
The following additional methods for evaluation of titration curves can be activated
and defined here:
Command name
25 characters, [ SET Upol # ]
Name of the command.
Fix end point evaluation
on, [ off ]
For a fixed quantity (measured value, volume or time) for the fix end point
the associated values are interpolated from the list of measuring points.
Fixed quantity
[ Measured value ], Time, Volume
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.
Fix EP# at
-200.0 ... 200.0 µA, [ off ]
Fix end point 1 ... 9 for
Fixed quantity
=
Measured value
.
Fix EP# at
0.0 ... 999999.9 s, [ off ]
Fix end point 1 ... 9 for
Fixed quantity
=
Time
.
Fix EP# at
0.00000 ... 9999.99 mL, [ off ]
Fix end point 1 ... 9 for
Fixed quantity
=
Volume
.
With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.
With this button instead of fix EP1...9 only the fix EP1...3 are shown.