3.2. Remote control commands

799 GPT Titrino

122

*Parameter

Tree part "Parameters for STAT"

.

CtrlPara

Control parameters

│ ├

.EP Measuring point

depends on meas.quant. 3.2.2.36.

│ ├

.UnitEp

Unit of endpoint

read only

ditto

│ ├

.Dyn

Dynamics

depends on meas.quant. 3.2.2.37.

│ ├

.UnitDyn

Unit of dynamics

read only

ditto

│ ├

.MaxRate

Maximum dosing rate

0.01...10...150, max.

ditto

│ ├

.MinRate

Minimum dosing rate

0.01...25.0...9999 3.2.2.40.

│ │

.

TitrPara

Titration parameters

│ ├

.StartV

Start volume

│ │ ├

.Type

Type of start volume

abs., rel., OFF 3.2.2.15.

│ │ ├

.V

Volume for absolute start volume

0...999.99 ditto

│ │ ├

.Factor

Factor for relative start volume

0...±999 999

ditto

│ │ ├

.Rate

Dispensing rate for start volume

0.01...150.0, max. ditto

│ ├

.Pause

Waiting time after start volume

0...999 999

3.2.2.16.

│ ├

.Tstart

Start time

0...999 999

3.2.2.41.

│ ├

.StartMeas

Start measuring value

depends on meas.quant.

ditto

│ ├

.RateStart Start

rate

0.01...150,

OFF ditto

│ ├

.TDelta

Time interv.for meas.acquisition

1...2...999 999

3.2.2.34.

│ ├

.Direction

Titration direction

+, -, auto 3.2.2.31.

│ ├

.DosUnit

Selection of the dosing unit

internal D0, external D1,

│ │

external

D2

3.2.2.17.

│ ├

.MeasInput Measuring

input

1, 2, diff.

3.2.2.18.

│ ├

.Ipol

Polarization

current

0...1...±127 ditto

│ ├

.Upol

Polarization voltage

0...400...±1270 ditto

│ ├

.PolElectrTest

Test for polarized electrodes

ON, OFF ditto

│ ├

.Temp

Titration temperature

-170.0...25.0...500.0 3.2.2.19.

│ │

.

StopCond

Stop conditions

│ ├

.TStop

Stop time

│ │ ├

.Type

Type of stop time

abs.,rel.,delta,delay,OFF 3.2.2.42.

│ │ ├

.Time

Time for absolute stop time

0...999 999 ditto

│ │ ├

.Factor

Factor for relative stop time

0...±999 999 ditto

│ │ ├

.Delta

Time after EP is once reached

0...999 999 ditto

│ │ ├

.Delay

Time after last dosing step

0...999 999 ditto

│ ├

.VStop

Stop volume

│ │ ├

.Type

Type of stop volume

abs., rel., OFF

3.2.2.20.

│ │ ├

.V

Volume for absolute stop volume

0...99.99...9999.99 ditto

│ │ ├

.Factor

Factor for relative stop volume

0...±999 999 ditto

│ ├

.RateStop Stop

rate

0.01...150.0,

OFF

3.2.2.43.

│ ├

.FillRate Filling

rate

0.01...150.0,

max. 3.2.2.23.

│ │

.

Statistics

Statistics

│ ├

.Status

Status of statistics calculation

ON, OFF 3.2.2.24.

│ ├

.MeanN

No. of individual determinations

2...20 ditto

│ ├

.ResTab Result

table

│ │ ├

.Select

original,delete n,delete all

ditto

│ │ ├

.DelN

Deletion of individual results

1...20 ditto

│ │