Metrohm 799 GPT Titrino Benutzerhandbuch

Seite 130

Advertising
background image

3.2. Remote control commands

799 GPT Titrino

126

*Parameter

Tree part "Parameters for DOC"

.

DosPara

Dosing parameters

│ ├

.BeginMeas

Begin of measuring

depends on meas.quant. 3.2.2.57.

│ ├

.EndMeas

End of measuring

depends on meas.quant.

ditto

│ ├

.UnitMeas

Unit of measured quantity

read only

ditto

│ ├

.SweepTime Sweep

time

0...300...999 999

3.2.2.58.

│ ├

.Dyn

Dynamics

depends on meas.quant. 3.2.2.59.

│ ├

.UnitDyn

Unit of dynamics

read only

ditto

│ ├

.MaxRate

Maximum dosing rate

0.01... 150, max. ditto

│ ├

.MinRate

Minimum dosing rate

0.01...5.0...9999 ditto

│ ├

.Direction

Direction +,

-,

auto 3.2.2.60.

│ ├

.StartV Start

volume

│ │ ├

.Type

Type of start volume

abs., rel., OFF 3.2.2.61.

│ │ ├

.V

Volume for absolute start volume

0...999.99 ditto

│ │ ├

.Factor

Factor for relative start volume

0...±999 999

ditto

│ │ ├

.Rate

Dispensing rate for start volume

0.01...150.0, max. ditto

│ ├

.Pause

Waiting time after start volume

0...999 999

ditto

│ ├

.TDelta

Time interv. for meas.acquisition

1...10...999 999

3.2.2.52.

│ ├

.DosUnit

Selection of the dosing unit

internal D0, external D1,

│ │

external

D2

3.2.2.17.

│ ├

.MeasInput Measuring

input

1, 2, diff.

3.2.2.62.

│ ├

.Ipol

Polarization

current

0...1...±127 ditto

│ ├

.Upol

Polarization voltage

0...400...±1270 ditto

│ ├

.PolElectrTest

Test for polarized electrodes

ON, OFF ditto

│ ├

.Temp

Titration temperature

-170.0...25.0...500.0 3.2.2.53.

│ │

.

StopCond

Stop conditions

│ ├

.VStop

Stop volume

│ │ ├

.Type

Type of stop volume

abs., rel., OFF

3.2.2.20.

│ │ ├

.V

Volume for absolute stop volume

0...99.99...9999.99 ditto

│ │ ├

.Factor

Factor for relative stop volume

0...±999 999 ditto

│ ├

.FillRate Filling

rate

0.01...150.0,

max. 3.2.2.23.

│ │

.

Statistics

Statistics

│ ├

.Status

Status of statistics calculation

ON, OFF 3.2.2.24.

│ ├

.MeanN

No. of individual determinations

2...20 ditto

│ ├

.ResTab Result

table

│ │ ├

.Select

original,delete n,delete all

ditto

│ │ ├

.DelN

Deletion of individual results

1...20 ditto

│ │

.

Monitoring

Monitoring

│ ├

.MeasVal

Monitoring of measured values

│ │ ├

.Status

Status

ON, OFF 3.2.2.45.

│ │ ├

.LowLim Lower

limit

depends on meas.quant.

ditto

│ │ ├

.UpLim Upper

limit

depends on meas.quant.

ditto

│ │ ├

.Action

Action if out of limits

end, hold, wait, none ditto

│ ├

.Temp

Monitoring of temperatures

│ │ ├

.Status

Status

ON, OFF 3.2.2.47.

│ │ ├

.LowLim Lower

limit -170.0...500.0 ditto

│ │ ├

.UpLim Upper

limit

-170.0...500.0 ditto

│ │ ├

.Action

Action if out of limits

end, hold, wait, none ditto