Metrohm 799 GPT Titrino Benutzerhandbuch

Seite 171

Advertising
background image

3.2 Remote control commands

799 GPT Titrino

167

Calibration data can be entered. They are accepted with &Info.CalibrationData

$G. If calibration data are entered, the calibration date is deleted.

3.2.2.132. Info.PrepData.D0.Date read

only

Info.PrepData.D0.Time read

only

Info.PrepData.D1.Type read

only

Info.PrepData.D1.Date read

only

Info.PrepData.D1.Time read

only

identical to .D2

Data from the last correctly executed preparation.

3.2.2.133. Info.Checksums $G

Info.Checksums.MPList read

only

Info.Checksums.ActualMethod read

only

The checksums can be used to identify the content of a file unequivocally,

e.g. files with identical content

have identical results of the checksums. An empty file has checksum "0". The

calculation of the checksums is triggered with $G.

.MPList: Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the

same results of the checksum.

3.2.2.134. Info.DetermData $G

Info.DetermData.Write ON,

OFF

Info.DetermData.MPList.1.Attribute read

only/read+write

Info.DetermData.ExV read

only/read+write

Info.DetermData.MPList.1.X read

only/read+write

Info.DetermData.MPList.1.Y read

only/read+write

Info.DetermData.MPList.1.Z1 read

only/read+write

Info.DetermData.MPList.1.Z2 read

only/read+write

for every measuring point

Determination data in hexadecimal format. A measuring point list is available

in mode DET, MET, STAT, DOS, DOC, SET, KFT, and MEAS.

Recalculation of the measuring data is triggered with $G.

.Write:

With "ON", the following nodes can be overwritten:

&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),

&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.

.ExV:

Volume of the exchange unit, with which the determination was

executed

.MPList.1.Attribute: Attribute

.MPList.X: X coordinate, time

.MPList.Y: Y coordinate, volume

.MPList.Z1: Z1 coordinate, measuring value

.MPList.Z2: Z2 coordinate, temperature

3.2.2.135. Info.TitrResults.RS.1.Value read

only

etc., up to .9

Info.TitrResults.EP.1.V read

only