3.2. Remote control commands

799 GPT Titrino

154

Mode.Parameter.Monitoring.MeasVal.pH.LowLim

0...±20.00

Mode.Parameter.Monitoring.MeasVal.pH.UpLim 0...±20.00

Mode.Parameter.Monitoring.MeasVal.U.LowLim

0...±2000

Mode.Parameter.Monitoring.MeasVal.U.UpLim 0...±2000

Mode.Parameter.Monitoring.MeasVal.Ipol.LowLim

0...±2000

Mode.Parameter.Monitoring.MeasVal.Ipol.UpLim 0...±2000

Mode.Parameter.Monitoring.MeasVal.Upol.LowLim

0...±200.0

Mode.Parameter.Monitoring.MeasVal.Upol.UpLim 0...±200.0

Mode.Parameter.Monitoring.MeasVal.Action end, hold, wait, none

Parameters for DOS: Selection of the measured quantity for measured value

monitoring. Measured values (pH, U and I) are entered in the list of measured

points only when the monitoring is active.

Selection of the measurement input (MeasInput) applies with measured quan-

tities pH and U. diff. means differential amplifier, see page 216.

With Ipol, the requests for the polarization current in uA (Ipol) and

.PolElectrTest apply.

With Upol, the request for the polarization voltage in mV (Upol) applies. Entry

in steps of 10 mV. .PolElectrTest also applies. If the test for polarized elec-

trodes is switched on (ON), it is performed on the change from the inactive

basic mode to an active mode.

On limit value violation, an action can be selected:

stop: Abort of the determination.

hold: Interrupt dosing until the sequence is continued by manual switching

(&m $C).

wait: Interrupt dosing until the limit value is no longer violated, then continue

automatically.


3.2.2.57. Mode.Parameter.DosPara.BeginMeas pH: 0...±20.00, init

U,

Ipol:

0...±2000, init

Upol:

0...±200.0, init

Mode.Parameter.DosPara.EndMeas

as with .Begin, but OFF instead of init

Mode.Parameter.DosPara.UnitMeas read

only

Parameters for DOS: Setting of ramp start and ramp end. init means ramp

start with default measured value.

3.2.2.58. Mode.Parameter.DosPara.SweepTime 0...300...999 999

Sweep time for the ramp in s.

3.2.2.59. Mode.Parameter.DosPara.Dyn pH:

0.01...0.25...20.00, OFF

U:

1...15...2000, OFF

Ipol:

1...6...2000, OFF

Upol:

0.1...10...200.0, OFF

Mode.Parameter.DosPara.UnitDyn read

only

Mode.Parameter.DosPara.MaxRate 0.01...150,

max.

Mode.Parameter.DosPara.MinRate 0.01...5.0...9999

Parameters for DOC: Control parameters, see page 68.

.Dyn: Dynamics in pH, mV (with U and Ipol) or uA (with Upol). The corre-

sponding unit can be read with .UnitDyn.