Metrohm 799 GPT Titrino Benutzerhandbuch

Seite 174

Advertising
background image

3.2. Remote control commands

799 GPT Titrino

170

3

Clear (pin 10)

3

EOD (pin 17)

4

Smpl Ready (pin 23)

4

Monitoring, line L4 (pin 3)

5

pin 11

5

Error (pin 16)

6

pin 24

6

Activate, line L6 (pin 1)

7

pin 12

7

Pulse for recorder (pin 2)

8

not used

(pin 6)

9

not used

(pin 7)

10

Monitoring (pin

8)

11

Monitoring (pin

13)

12

Monitoring (pin

19)

13

Monitoring (pin

20)

3.2.2.141. Info.ActualInfo.Assembly.CyclNo read

only

$Q sends the current cycle number of the voltage measurement cycle, e.g.

"127". From the cycle number and the cycle time (see 3.2.2.149), a time

frame can be set up.

The cycle number is set to 0 on switching on the instrument, on every start

and for QuickMeas. It is incremented as long as the instrument remains

switched on.

3.2.2.142. Info.ActualInfo.Assembly.Counter.V read

only

Info.ActualInfo.Assembly.Counter.Clear $G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the

volume counter is set to zero.

3.2.2.143. Info.ActualInfo.Assembly.Meas read

only

$Q sends the current measured value from the assembly.

3.2.2.144. Info.ActualInfo.Titrator.CyclNo read

only

Info.ActualInfo.Titrator.V read

only

Info.ActualInfo.Titrator.Meas read

only

Info.ActualInfo.Titrator.dVdt read

only

Info.ActualInfo.Titrator.dMeasdt read

only

Info.ActualInfo.Titrator.dMeasdV read

only

Info.ActualInfo.Titrator.ERC read

only

Info.ActualInfo.Titrator.T read

only

$Q sends the current values in the following formats:

DET

MET

SET

KFT

STAT,

DOC

DOS MEAS

CAL

CyclNo

127 127 127 127 127 127 127 127

V(ml)

1.2345 1.2345 1.2345 1.2345 1.2345 1.2345 -

-

Meas:

pH

U, Ipol (mV)

Upol (uA)

T (°C)

3.345

-345.6

-12.5

-

3.345

-345.6

-12.5

-

3.6(mV)

-345.6

-12.5

-

-

-345.6

-12.5

-

3.6(mV)

-345.6

-12.5

-

3.6(mV)

-345.6

-12.5

3.345

-345.6

-12.5

25.0

3.345

-

-

-

dVdt

(ul/s)

-

-

2.5142 2.5142 2.5142 -

-

-

dMeasdt