Metrohm 799 GPT Titrino Benutzerhandbuch

Seite 150

Advertising
background image

3.2. Remote control commands

799 GPT Titrino

146

If an absolute start volume (abs.) has been selected, the volume in mL is

valid.

A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size

*

factor

The factor is valid.

The dispensing rate in mL/min applies to both cases. Max. means maximum

possible dispensing rate with the Exchange Unit in current use.

3.2.2.16. Mode.Parameter.TitrPara.Pause

0...999 999

Parameters for DET, MET, SET, KFT, STAT: Pause time in s. Is waited off after

the dispensing of the start volume.

3.2.2.17. Mode.Parameter.TitrPara.DosUnit internal

D0, external D1/D2

Mode.Parameter.DosPara.DosUnit internal

D0, external D1/D2

.TitrPara.DosUnit: Parameter for DET, MET, SET, KFT, STAT. Selection of dos-

ing unit.

.DosPara.DosUnit: Parameter for DOS, DOC. Selection of dosing unit.

3.2.2.18. Mode.Parameter.TitrPara.MeasInput

1, 2, diff.

Mode.Parameter.TitrPara.Ipol -127...1...+127

Mode.Parameter.TitrPara.Upol -1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest ON, OFF

Parameters for DET, MET, SET, KFT STAT:

Selection of the measuring input; valid with measured quantities pH and U.

"diff." means differential amplifier, see page 215.

With Ipol, the inquiries for the polarization current in uA (Ipol) and

.PolElectrTest are valid.

With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in

steps of 10 mV.

Besides .PolElectrTest is valid.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to an active state (titration or conditioning).

3.2.2.19. Mode.Parameter.TitrPara.Temp -170.0...25.0...500.0

Parameters for DET, MET, SET, KFT, STAT: Titration temperature in °C. If a

Pt100 or Pt1000 is connected, the temperature is measured continuously and

the parameter .Temp is updated.

The temperature is used for the temperature correction in pH measurements.

3.2.2.20. Mode.Parameter.StopCond.VStop.Type abs., rel., OFF

Mode.Parameter.StopCond.VStop.V 0...99.99...9999.99

Mode.Parameter.StopCond.VStop.Factor 0...±999 999

Parameters for DET, MET, SET, KFT, STAT, DOS, DOC: Stop volume. (With

DOS the default value for .Type is "OFF" and the input range for V is

0...99999.99.)

If an absolute stop volume (abs.) has been selected, the volume in mL is

valid.

A relative stop volume (rel.) is dispensed as a function of the sample size: